Views: 222 Author: Dream Publish Time: 2025-04-23 Origin: Site
Content Menu
● Understanding Nobel Active Abutment Screw Torque
● Tools Needed to Measure Torque
● Step-by-Step Guide to Measuring Torque for Nobel Active Abutment Screw
>> Step 2: Assemble the Torque Wrench
>> Step 3: Hand Tighten the Abutment Screw
>> Step 4: Apply Torque with the Wrench
>> Step 5: Retighten After 10 Minutes
● Important Tips for Accurate Torque Measurement
● Troubleshooting Common Issues
>> Abutment Loose but Screw Tight
>> Screw Stripping or Breaking
>> Difficulty Achieving Recommended Torque
● The Role of Torque in Long-Term Implant Success
● Advanced Considerations: Digital Torque Measurement and New Technologies
● FAQ
>> 1. What is the recommended torque for a Nobel Active abutment screw?
>> 2. How do I ensure accurate torque measurement with a Nobel Biocare torque wrench?
>> 3. What should I do if the abutment feels loose but the screw is tight?
>> 4. Can I use a hand screwdriver to tighten Nobel Active abutment screws?
>> 5. How often should I retighten the abutment screw after initial torque application?
Measuring the correct torque for a Nobel Active abutment screw is critical for the success and longevity of dental implant restorations. Proper torque ensures the stability of the abutment, prevents screw loosening, and maintains the integrity of the implant system. This comprehensive guide will walk you through the process of measuring torque for a Nobel Active abutment screw, including the tools required, step-by-step instructions, tips for accuracy, and troubleshooting common issues. We will also discuss the recommended torque values and the importance of following manufacturer guidelines.
The Nobel Active implant system is designed for high primary stability and versatility in various bone qualities. The abutment screw torque for Nobel Active implants is a specific value recommended by the manufacturer to ensure optimal preload and retention without damaging the components.
Recommended Torque for Nobel Active Abutment Screw:
The clinical consensus and manufacturer guidelines recommend tightening the Nobel Active abutment screw to 15 Ncm (Newton centimeters) for prosthetic screws, while some final abutments may require up to 35 Ncm depending on the abutment type and clinical situation.
Why Torque Matters:
Applying the correct torque ensures the abutment is securely fastened, reducing micro-movement and preventing screw loosening, which can lead to prosthetic failure or biological complications.
The Science Behind Torque and Preload:
Torque applied to the abutment screw generates a preload, which is the tension within the screw that holds the abutment firmly against the implant. Adequate preload is essential to resist functional forces during chewing and prevent screw loosening or fracture. Too low torque results in insufficient preload, while too high torque risks screw or implant damage.
To accurately measure torque for a Nobel Active abutment screw, you will need:
Nobel Biocare Manual Torque Wrench:
A calibrated torque wrench designed for dental implants, specifically compatible with Nobel Active components. These wrenches provide precise torque control and a tactile or audible signal (click or ratchet) when the preset torque is reached.
Unigrip Screwdriver or Torque Driver:
Used to engage the abutment screw head securely before applying torque. The driver must fit perfectly to avoid cam-out or damage to the screw head.
Torque Calibration and Maintenance:
Torque wrenches must be regularly cleaned, sterilized, and recalibrated to maintain accuracy. Improper maintenance can lead to incorrect torque application, risking screw damage or loosening.
Additional Accessories:
Depending on the clinical setup, you may also require implant analogs, impression copings, or verification jigs to confirm the accuracy of the abutment seating before torque application.
Verify Components:
Ensure you are using original Nobel Biocare abutment screws and compatible drivers to avoid mismatches that can affect torque accuracy.
Clean and Sterilize Tools:
Clean and autoclave the torque wrench and drivers as per manufacturer instructions to maintain hygiene and function.
Inspect the Implant and Abutment:
Before assembly, inspect the implant platform and abutment connection for any debris, damage, or irregularities that could affect seating or torque.
- Attach the appropriate driver (short, medium, or long) to the torque wrench.
- Confirm the directional arrow on the torque wrench is facing toward the operator to ensure proper torque application.
- Begin by hand-tightening the abutment screw into the implant to ensure it is properly seated and aligned. This step reduces the risk of cross-threading and ensures the screw head is flush with the abutment surface.
- Place the torque wrench on the driver and slowly apply force until the wrench clicks or ratchets, indicating the preset torque value has been reached.
- For Nobel Active abutment screws, torque to 15 Ncm for prosthetic screws or 35 Ncm for final abutments as specified.
- Note: Apply torque in a slow, steady motion to avoid sudden force spikes that could damage the screw or implant.
- It is recommended to retighten the screw after 10 minutes to compensate for any initial settling or preload loss. This practice helps maintain the correct torque over time and reduces the risk of screw loosening.
Use the Correct Driver:
Using the wrong driver can cause the wrench to slip and fail to apply the correct torque. Always use the driver supplied or recommended by Nobel Biocare for the specific abutment screw.
Avoid Over-Tightening:
Exceeding the recommended torque can damage the implant or surrounding bone. Nobel Biocare recommends not exceeding 70 Ncm insertion torque for implants, with abutment screws at much lower values (15-35 Ncm).
Calibrate Torque Wrench Regularly:
Sterilization and use can affect wrench accuracy; regular calibration is essential. Follow manufacturer guidelines for calibration intervals, typically every 6-12 months or after a set number of uses.
Check for Debris:
Clean the screw threads and implant connection to avoid torque inaccuracies caused by debris or contamination. Use sterile saline or alcohol wipes to clean the surfaces before assembly.
Apply Torque in a Controlled Environment:
Avoid applying torque in awkward positions or under excessive patient movement, which can cause inaccurate torque application.
Document Torque Values:
Record the applied torque in patient records for quality control and future reference.
This can occur if the abutment is not fully seated into the implant hex, causing rotation despite a tight screw. Re-seat the abutment or replace the screw if necessary. Also, check for any deformation or damage to the implant connection.
Usually due to excessive torque or damaged screws. Replace screws and adhere strictly to recommended torque values. Use only original Nobel Biocare screws, as third-party components may have different tolerances.
Check that the directional arrow is correctly aligned and the knurled knob is tightened to ensure proper function. If the wrench continues to malfunction, send it for recalibration or replacement.
If resistance is unusually high or low, inspect the screw threads and implant interface for contamination or damage. Lubrication is generally not recommended unless specified by the manufacturer.
Correct torque application is not just a procedural step; it directly impacts the longevity and success of the implant restoration. Studies have shown that inadequate torque can lead to micro-movements at the implant-abutment interface, causing screw loosening, peri-implant bone loss, and eventual implant failure. Conversely, over-torquing can induce stress fractures in the screw or implant body.
Nobel Active implants are designed to maximize primary stability, especially in challenging bone conditions such as soft or low-density bone. Proper torque application complements this design by ensuring the abutment remains stable under functional loads.
Recent advances in dental technology have introduced digital torque measurement devices that provide real-time feedback and precise torque control. These devices can be integrated with implant surgical guides and digital workflows for enhanced accuracy.
Digital Torque Drivers:
Provide visual and audible alerts when the preset torque is reached, reducing operator variability.
Smart Torque Wrenches:
Some models can record torque data and transmit it wirelessly for documentation and quality assurance.
While these technologies are promising, the Nobel Biocare manual torque wrench remains the clinical standard due to its reliability, ease of sterilization, and proven accuracy.
Measuring the torque for a Nobel Active abutment screw is a precise process that requires the correct tools, adherence to manufacturer torque specifications, and careful technique. The recommended torque values are typically 15 Ncm for prosthetic screws and up to 35 Ncm for final abutments, with a strict upper limit to prevent damage. Using a calibrated Nobel Biocare torque wrench with the correct driver ensures the screw is tightened properly, maintaining implant stability and longevity. Retightening after 10 minutes and regular maintenance of the torque wrench further enhance clinical outcomes. Understanding and applying these principles is essential for implantologists and restorative dentists working with Nobel Active systems.
The recommended torque for a Nobel Active abutment screw is typically 15 Ncm for prosthetic screws and up to 35 Ncm for final abutments, depending on the clinical situation and abutment type.
Ensure the torque wrench is properly assembled with the correct driver, the directional arrow is facing the operator, the knurled knob is tightened, and the wrench is regularly calibrated and sterilized according to manufacturer instructions.
This may indicate the abutment is not fully seated into the implant hex or there is debris in the threads. Try reseating the abutment, cleaning the connection, or replacing the screw if necessary.
No, hand screwdrivers do not provide the precise torque needed. A calibrated torque wrench is mandatory to apply the correct torque and avoid complications.
It is recommended to retighten the abutment screw after 10 minutes to compensate for preload loss and ensure stable torque retention.
[1] https://adldental.com/wp-content/uploads/2018/10/ADL_Torque-Spec-Sheet3.pdf
[2] https://www.youtube.com/watch?v=leUA1mqve8c
[3] https://www.straumann.com/content/dam/media-center/medentika/en/documents/brochure/product-information/PM02_01_0044_EN_F-Serie_print.pdf
[4] https://store.nobelbiocare.com/media/eifu/IFU1020_EN_GB_02.pdf
[5] https://www.youtube.com/watch?v=VNkmT2jjpG0
[6] https://www.youtube.com/watch?v=MnSjPrJaLvY
[7] https://www.spotimplant.com/en/knowledge-base/abutment-screw-torque-and-driver/
[8] https://www.osseonews.com/abutment-loose-but-screw-tight/
[9] https://meridian.allenpress.com/joi/article/35/6/283/2260/Clinical-Evaluation-of-the-NobelActive-Implant
[10] https://www.novodent.hr/sites/default/files/field/file/proizvod/79424_guided_surgery_with_nobelactive_manual_2015_gb.pdf
[11] https://alamoimplantcenter.com/wp-content/uploads/2020/01/Abutment-Guide.pdf
[12] https://blog.iti.org/clinical-insights/angulated-screw-channel-solutions-in-implant-dentistry/
[13] https://journals.lww.com/jorr/fulltext/2018/10020/strategies_to_reduce_torque_loss_of_abutment.4.aspx
[14] https://www.burdettedental.com/wp-content/uploads/2016/11/7204-Torque-Spec-Sheet.V5.web_.pdf
[15] https://www.nobelbiocare.com/sites/g/files/wdvifx201/files/Prosthetic%20Torque%20Guide.pdf
[16] https://www.nobelbiocare.com/en-int/top-questions-frequently-asked-questions-faqs
[17] https://edeldental.hu/images/pdf/nobelactive_sales_brochure_en.pdf
[18] https://www.helmdentallaboratory.com/tpn/c/C1050/docs/20200713-c1050-Helm-Dental-Torque-Values.pdf
[19] https://alamoimplantcenter.com/wp-content/uploads/2020/01/Abutment-Guide.pdf
[20] https://www.eotdental.com/uploads/pdf/EOT_TR_PRODUCT_BOOK.pdf
[21] https://c1-preview.prosites.com/65806/wy/docs/NobelActive_Product_and_Procedures_GB_tcm57-13044.pdf
[22] https://store.nobelbiocare.com/media/eifu/IFU1020_EN_GB_02.pdf
[23] https://renstrom.squarespace.com/s/Implant-Torque-Guide.pdf
[24] https://triaddentalstudio.com/wp-content/uploads/2019/09/Updated_Implant-Abutment-Torque-Settings-Resized-to-Letter.pdf
[25] https://www.osseonews.com/abutment-loose-but-screw-tight/
[26] https://www.dess-usa.com/nobel-active-compatible-screws/
[27] https://store.nobelbiocare.com/us/en/guided-template-abutment-with-screw-nobelactive-np-1
[28] https://www.youtube.com/watch?v=5HvB8i3z5x0
[29] https://www.youtube.com/watch?v=rJRxl504vgU
[30] https://www.dentalkart.com/mediden-nobel-active-np-titanium-abutment-screw-s-46.html
[31] https://www.youtube.com/watch?v=x_SNgKu3Clw
[32] https://www.nobelbiocare.com/en-int/system/files/gmt_import/Guided%20Surgery%20Manual_87724A.pdf
[33] https://www.youtube.com/watch?v=-jUPqZkjklM
[34] https://www.nobelbiocare.com/en-int/technical-support-frequently-asked-questions-faqs
[35] https://www.youtube.com/watch?v=oCH0ii_r3Os
[36] https://www.burdettedental.com/wp-content/uploads/2016/11/7204-Torque-Spec-Sheet.V5.web_.pdf
[37] https://www.eotdental.com/uploads/pdf/NobelParallel-Manual%20.pdf
[38] https://www.nobelbiocare.com/en-in/top-questions-frequently-asked-questions-faqs
[39] https://adldental.com/wp-content/uploads/2018/10/ADL_Torque-Spec-Sheet3.pdf
[40] https://www.perioimplantadvisory.com/dental-implants/article/14293534/dental-implant-torque-wrench-error-solving-the-problem
[41] https://www.nobelbiocare.com/en-int/angulated-screw-channel-solutions
[42] https://rdl-usa.com/support/dental-implant-torque-values-guide/
[43] https://www.straumann.com/content/dam/media-center/medentika/en/documents/brochure/product-information/PM02_01_0044_EN_F-Serie_print.pdf
[44] https://dpsdental.com/wp-content/uploads/2022/04/Implant-Torque-Guide-2022.pdf
[45] https://pmc.ncbi.nlm.nih.gov/articles/PMC3607726/
[46] https://glidewelldental.com/content/dam/digital/websites/document-library/GD-853.pdf.librarydownload.inline.pdf
What Are The Common Complications From Stripped Abutment Screws?
What Is The Difference Between Titanium And Gold Straumann Abutment Screws?
Are Straumann Abutment Torque Values Different For Custom And Standard Abutments?
How Effective Is The Straumann Screw Removal Kit For Dental Implants?
Can Stainless Steel Abutment Screws Be Used In All Implant Systems?
How Do You Replace A Short Abutment Screw In Hiossen Implants?